Low-Temperature Impact Test Device MLT-LTI 901

Product information:
The HT Apparatus for impact test at low temperature confirms to the standard of IEC 60884 figure 27 and figure 42. The first one is fit for impact test of wire while the latter one is applied for the impact test of plug, socket, switch, wire and electrical conduit.

Technical parameters:

This device is used with the cryostat. It should be put in the sponge-rubber pad of 40mm-thickness. The height of falling: 10~250mm is adjustable.

Low-temperature impact test sample standard Weight of the drop hammer (g) the height of the impact (mm)
The socket box and the similar electrical enclosure IEC 60884-1 clause24.4, figure 27 1000±2 100
The bolt  with insulation sheath IEC 60884-1  clause30.4, figure 42 100±1 100
The electric cable with PVC sheath IEC 60884-1  clause8.5,figure2 100、200、300、 100
400、500、600、
750 、 1000 、
1250、1500g


For the samples of laying of cable, the weight of the drop hammer required are as follows:

The external diameter of the samples(D/mm) The weight of the drop hammer /g
D≤4.0 100
4.0<D≤6.0 200
6.0<D≤9.0 300
9.0<D≤12.5 400
12.5<D≤20.0 500
20.0<D≤30.0 750
30.0<D≤50.0 1000
50.0<D≤75.0 1250
75.0<D 1500


For the samples of flexible cable and telecommunication cable, the weight of the drop hammer required is as follows:

The external diameter of the sample D/mm The weight of drop hammer /g
D≤6 100
6.0<D≤10.0 200
10.0<D≤15.0 300
15.0<D≤25.0 400
25.0<D≤35.0 500
35.0<D 600

 

Low-Temperature Impact Test Device MLT-LTI 901